Leave Your Message
Products Categories
Featured Products

65W Longxc 65W Single Output Medical Grade series 5V 12V 24V longxc power ACMS59

The purpose of the document is to specify the functional requirements of a ACMS59 series medical adapter.

    Input Characteristics

    Input Voltage:

    Nominal Voltage: 100~240Vac

    Variation Range: 90~264Vac

    Input Frequency:

    Nominal Frequency: 50/60Hz.

    Variation Frequency: 47~63Hz

    Input Current:

    1.4A max. at 90Vac input and rated load.

    Inrush Current:

    Output rated load and ambient temperature 25℃,maximum inrush current less than 80A after 5µs at 264Vac input cold start.

    AC Leakage Current:

    Normal touch current less than 0.1mA at 264Vac input.

    Normal earth leakage current less than 5mA at 264Vac input.

    Output Characteristics

    Model name

    Output voltage(V)

    Output range(V)

    Rated load(A)

    Output power (W)

    Ripple and noise(mVp-p)

    ACMS59-090

    9.0

    8.73~ 9.27

    0-5.50

    49.5

    120

    ACMS59-120

    12.0

    11.64~12.36

    0-5.50

    66

    120

    ACMS59-135

    13.5

    13.10~13.90

    0-4.81

    65

    135

    ACMS59-142

    14.2

    13.77~14.63

    0-4.58

    65

    142

    ACMS59-150

    15.0

    14.55~15.45

    0-4.33

    65

    150

    ACMS59-168

    16.8

    16.30~17.30

    0-3.87

    65

    168

    ACMS59-180

    18.0

    17.46~18.54

    0-3.61

    65

    180

    ACMS59-190

    19.0

    18.43~19.57

    0-3.42

    65

    190

    ACMS59-200

    20.0

    19.40~20.60

    0-3.25

    65

    200

    ACMS59-220

    22.0

    21.34~22.66

    0-2.95

    65

    220

    ACMS59-240

    24.0

    23.28~24.72

    0-2.71

    65

    240

    ACMS59-260

    26.0

    25.22~26.78

    0-2.50

    65

    260

    ACMS59-280

    28.0

    27.16~28.84

    0-2.32

    65

    280

    ACMS59-300

    30.0

    29.10~30.90

    0-2.17

    65

    300

    ACMS59-320

    32.0

    31.04~32.96

    0-2.03

    65

    320

    ACMS59-340

    34.0

    32.98~35.02

    0-1.91

    65

    340

    ACMS59-360

    36.0

    +

    34.92~37.08

    0-1.81

    65

    360

    ACMS59-380

    38.0

    36.86~39.14

    0-1.71

    65

    380

    ACMS59-400

    40.0

    38.80~41.20

    0-1.63

    65

    400

    ACMS59-420

    42.0

    40.74~43.26

    0-1.55

    65

    420

    ACMS59-440

    44.0

    42.68~45.32

    0-1.48

    65

    440

    ACMS59-460

    46.0

    44.62~47.38

    0-1.41

    65

    460

    ACMS59-480

    48.0

    46.56~49.44

    0-1.35

    65

    480

    Line regulation

    ±3%

    Load regulation

    ±3%

    Note 1

    Test conditions: under nominal voltage and rated load,bandwidth of 20MHz and parallel 10uF/0.1uF, connected at testing point.

    Ripple and Noise

    Test conditions: Under nominal voltage and rated load, the ripple and noise are less than 350mVp-p when measure with Max. Bandwidth of 20MHz and parallel 10uF/0.1uF, connected at testing point. 

    Turn on delay time

    3 second Max. at 90Vac input and output rated load.

    Rise time

    40ms Max. test condition when output rated load.

    Hold up time

    10ms min. at 90Vac input and output rated load.

    Efficiency:

    83% min. at 220Vac input voltage,rated load.

    Protection Function

    Short circuit test

    The power supply will be auto recovering when short circuit faults removed.

    Over current Protection

    When the output current reaches 110%~200% of the rated current, the power supply will be protected. After the overcurrent fault is removed, the power supply will be automatically recovery.

    Over Voltage Protection

    When the output voltage reaches 105%~140% of the rated voltage,the power supply will be protected and can resume normal work after the fault removed.

    Electromagnetic Compatiblity

    Numbe

    Item

    Specifications

    Class

    Standards

    1

    (CE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    2

    (RE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    3

    (SURGE)

    Line to line±1KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-5; GB17626.5

     

     

    Line to GND±2KV

    A

     

    4

    (ESD)

    Air discharge±15KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-2; GB17626.2

     

     

    Contact discharge±8KV

    A

     

    5

    (EFT/B)

    ±2KV (BURST FREQUENCY=100KHZ)

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-4; GB17626.4

    6

    (DIP)

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-11; GB17626.11

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

    Dropped to0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    (RS)

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;80%AM(1KHz)

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-3; GB17626.3

    8

    (CS)

    Test frequency:0.15MHz~80MHz;

    Field intensity:6Vrms;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-6; GB17626.6

    9

    (THD)

    CLASSA(in system)

    /

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-3-2; GB17625.1

    10

    Voltage

    Fluctuation and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-3-3; GB17625.2

    11

    Power Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-8; GB17626.8

    Performance criterion A: Normal performance within limits in this specification. 
    Performance criterion B: Temporary loss of some functions or degradation of performance. The recovering performance of which without operator intervention.
    Performance criterion C: Temporary loss of function or degradation of performance .The recovering performance of which requires operator intervention.
    Performance criterion D: Loss of function or degradation of performance which is not recoverable .Owing to damage of hardware software, or loss of data.

    Safety: accord with

    The power supply is designed for indoor use to meet IEC 60601-1/UL60601-1and EN 60601-1standards and for the following safety regulations.
     

    Item

    Country

    Standard

    UL

    USA

    UL60601-1

    CE

    Europe

    EN60601-1

    CB

    Worldwide

    IEC60601-1

    TUV

    Germany

    IEC60601-1

    NRTL

    CUSA

    IEC60601-1/UL60601-1

    GS

    Germany

    EN60601-1

    BS

    England

    EN60601-1

    SAA

    Australia

    AS/NZS6-1

    The primary to secondary: 4000Vac 10mA for 60 seconds
    The primary to earth: 1500Vac 5mA for 60 seconds
    The secondary to earth:500Vdc 5mA for 60 seconds

    Mechanical Requirement

    EU ping the power supply size: L 78.0 x W 35.6 x H 24.1mm;
    US ping the power supply size: L 40.5 x W 34.7 x H 24.1mm;
    asdasdsd 4mps

    Label

    Label size:30*20(mm)

    5cm2rwh