Leave Your Message
Products Categories
Featured Products

36W Hot sale model LXCP36 12V 3A medical AC/DC Power adapters for infusion pump LXCP36

The document aims to outline the functional requirements of a LXCP37 series switching power supply.

    Input Characteristics

    Input Voltage:

    Nominal Voltage: 100~240Vac

    Variation Range: 90~264Vac

    Input Frequency:

    Nominal Frequency: 50/60Hz.

    Variation Frequency: 47~63Hz

    Input Current:

    0.8Amps max at any input voltage and rated DC output and rated load.

    Inrush Current:

    60Amps Max. Cold start at 264Vac input, with rated load and 25℃ ambient.

    AC Leakage Current:

    Normal 0.5mA Max.at 264Vac input.

    Single fault 1.0mA Max.at 264Vac input.

    Output Characteristics

    Model name

    Output Voltage(V)

    Rated Load

    (A)

    Output Range(V)

    Over current protection(A)

    Efficiency

    (%)

    Output power(W)

    LXCP36-050

    5.0

    4.00

    4.40~5.60

    ≤8.0

    70

    20.0

    LXCP36-075

    7.5

    4.00

    6.90~8.10

    ≤8.0

    70

    30.0

    LXCP36-090

    9.0

    4.00

    8.40~9.60

    ≤8.0

    72

    36.0

    LXCP36-120

    12.0

    3.00

    11.40~12.60

    ≤6.0

    80

    36.0

    LXCP36-135

    13.5

    2.66

    12.90~14.10

    ≤5.0

    80

    36.0

    LXCP36-150

    15.0

    2.40

    14.40~15.60

    ≤5.0

    80

    36.0

    LXCP36-168

    16.8

    2.14

    16.20~17.40

    ≤4.0

    80

    36.0

    LXCP36-180

    18.0

    2.00

    17.40~18.60

    ≤4.0

    80

    36.0

    LXCP36-190

    19.0

    1.89

    18.40~19.60

    ≤3.6

    80

    36.0

    LXCP36-200

    20.0

    1.80

    19.40~20.60

    ≤3.6

    80

    36.0

    LXCP36-240

    24.0

    1.50

    23.40~24.60

    ≤3.0

    80

    36.0

    Line Regulation

    ±3%

    Load Regulation

    ±8%

    Ripple and Noise

    At nominal voltage and rated load, the ripple and noise measure less than 200mVp-p when using a maximum bandwidth of 20MHz and connecting a parallel 10uF/0.1uF at the testing point.

    Turn on delay time

    3 Second Max.at 220Vac input and output rated load.

    Rise time

    50mS Max.at 115Vac input and output rated load.

    Hold up time

    5mS Min.at 115Vac input and output rated Load.

    Protection Function

    Short circuit test

    The power supply will be auto recovering when the short circuit faults removed.

    Over current Protection

    The power supply will enter a stop mode if the IC's temperature exceeds its trigger point. It will automatically recover when the IC temperature falls below the regulation value.

    Environmental Requiremet

    Operating Temperature

    0℃ to 40℃,Rated load, Normal operation.

    Storage Temperature:-20℃ to 80℃

    With package

    Storage Humidity: 10%~90%

    With Package

    Atmospheric pressure

    70-106KPa, Normal.

    Altitude

    5000m,Work temp descend 1℃ every 300m over 5000m.

    (9~200Hz,Acceleration 5m/S2)

    Transportation: 5-9Hz, A=3.5mm

    Acceleration=5m/S2

    Acceleration=15m/S2

    Axes, 10 cycles per axis

    Testing should not result in any permanent damage.

    The power supply can be restored to normal operation by power cycling.

    Dropping Packed

    As described above, the wall mount type requires a 1m distance, while the desktop type requires a 760mm distance.

    The horizontal surface should be constructed with hardwood at least 13mm thick, mounted on two layers of plywood, and positioned 19mm to 20mm away from the edge.

    Relative Humidity

    5%(0℃) ~90%(40℃)RH, 72Hrs, Rated load, Normal operating.

    MTBF

    The power supply is expected to have a minimum predicted MTBF (MIL-STD-217F) of 100,000 hours under normal operating conditions and typical usage.

    Electromagnetic Compatiblity

    Number

    Item

    Specifications

    Class

    Standards

    1

    CE

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    2

    RE

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    3

    SURGE

    Line to line±1KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-5; GB17626.5

    Line to GND±2KV

    A

     

    4

    ESD

    Air discharge±15KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-2; GB17626.2

    Contact discharge±8KV

    A

     

    5

    EFT/B

    ±2KV (BURST FREQUENCY=100KHZ)

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-4; GB17626.4

    6

    DIP

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-11; GB17626.11

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

    Dropped to0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    RS

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-3; GB17626.3

    8

    CS

    Test frequency:0.15MHz~80MHz;

    Field intensity:6Vrms;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-6; GB17626.6

    9

    THD

    CLASSA(in system)

    /

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-3-2; GB17625.1

    10

    Voltage

    Fluctuation and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-3-3; GB17625.2

    11

    Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-8; GB17626.8

    Safety: accord with

    The power supply is designed for indoor use to meet IEC 60601-1 and EN 60601-1standards and for the following safety regulations.
     

    ITEM

    COUNTRY

    STANDARD

    UL

    USA

    UL60950-1/UL60601-1

    CE

    EUROPE

    EN60950-1/EN60601-1

    CB

    WORLDWIDE

    IEC60601-1

    TUV

    GERMANY

    IEC60601-1

    NRTL

    CUSA

    IEC60601-1/UL60601-1

    GS

    GERMANY

    EN60601-1

    BS

    ENGLAND

    EN60601-1

    SAA

    AUSTRALIA

    AS/NZS6-1

    The primary to secondary: 4000Vac 10mA for 60 seconds

    Mechanical Requirement

    AUS the power supply size: L 80.0 x W44.0 x H 56.7mm;
    GUSIGHDKHFLJ (1)t55
    EU the power supply size: L 80.0 x W 44.0 x H 67.9mm;
    GUSIGHDKHFLJ (2)j2y
    CN the power supply size: L 80.0 x W44.0 x H 53.3mm;
    GUSIGHDKHFLJ (3)r4c
    US the power supply size: L 80.0 x W44.0 x H 53.3mm;
    GUSIGHDKHFLJ (4)jw0
    Indian the power supply size: L 80.0 x W44.0 x H 56.1mm;
    GUSIGHDKHFLJ (5)8ar

    Cable

    detachable the power supply size: L 80.0 x W44.0 x H 38.1mm;

    ABUOxms

    optional DC plug 

    optional DC plug  (1)skdoptional DC plug  (2)j9doptional DC plug  (3)k9yoptional DC plug  (4)1kd
    Note: Details are selected from the LXC_ Wire Library drawing XLS file

    label size:34mm*26mm

    13qps4y1u

    FAQ

    What type of medical power supply does Longxc mainly produce? 
    From 1-600W power supply. Our medical power supply cover monitoring, anesthesia, respiratory, electrocardiogram machine, infusion pump, injection pump, B-ultrasound, imaging, biochemistry, electronic blood pressure monitor, beauty, rehabilitation physiotherapy and other medical equipment
    The development history of longxc? 
    Shenzhen Longxc Power Supply Co., Ltd. is a professional high-tech enterprises that engaged in medical switching power supply + safety EMC value-added services. Our company was founded in 2009, has a professional design and rectification team to provide medical equipment medical power solutions, focusing on the medical power supply industry more than 15 years. Safely escort for a total of 300 million units of medical equipment