24W US EU UK Plug Medical Grade 12V 2A Wall Mount IEC 60601 Brest Pump Adapter LXCP26

This document introduces the electrical, mechanical and environmental specifications of a switching power supply.

    INPUT CHARACTERISTICS

    INPUT VOLTAGE & FREQUENCY

    Minimum

    Nominal

    Maximum

    Input Voltage

    90Vac

    100-240Vac

    264Vac

    Input Frequency

    47Hz

    50Hz/60Hz

    63Hz

    INPUT CURRENT

    The maximum input current is 0.8A at nominal AC input voltage and full load.

    INRUSH CURRENT

    The inrush current will not exceed 50A at 264Vac input for a cold start at 25℃.

    STATIC LOSS 

    The static loss is less than 0.5W at normal input voltage.

    The efficiency shall be higher than 70% while measuring at 115VAC&230VAC input voltage, full load.

    OUTPUT CHARACTERISTICS

    Item

    Model number

    OUTPUT CHARACTERISTICS

    LXCP26series

    Rated output voltage(V)

    Output voltage range(V)

    Rated output current(A)

    Rated output power(W)

    Model number

    LXCP26-050

    5.0

    4.40~5.60

    3.00

    15.0

    LXCP26-060

    6.0

    5.40~6.60

    2.50

    15.0

    LXCP26-075

    7.5

    7.00~8.10

    2.66

    20.0

    LXCP26-090

    9.0

    8.50~9.70

    2.22

    20.0

    LXCP26-120

    12.0

    11.30~12.70

    2.00

    24.0

    LXCP26-126

    12.6

    11.90~13.30

    1.90

    24.0

    LXCP26-135

    13.5

    12.80~14.20

    1.77

    24.0

    LXCP26-150

    15.0

    14.30~15.70

    1.60

    24.0

    LXCP26-168

    16.8

    16.10~17.50

    1.43

    24.0

    LXCP26-180

    18.0

    17.30~18.70

    1.35

    24.3

    LXCP26-190

    19.0

    18.30~19.70

    1.26

    24.0

    LXCP26-200

    20.0

    19.30~20.70

    1.20

    24.0

    LXCP26-240

    24.0

    23.30~24.70

    1.00

    24.0

    LXCP26-260

    26.0

    25.30~26.70

    0.92

    24.0

    LXCP26-280

    28.0

    27.30~28.70

    0.85

    24.0

    LXCP26-300

    30.0

    29.30~30.70

    0.80

    24.0

    LXCP26-320

    32.0

    31.30~32.70

    0.80

    25.6

    load regulation

    ±5%

    voltage regulation

    ±3%

    Overshoot

    ≤5%

    Hold up time

    ≥10 mS

    Efficiency

    ≥70%

    Over current protection(A)

    110-150%Iout

    Ripple and noise

    ≤200mV (measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a10uF aluminum electrolytic capacitor)

    Over temperature protection

    The power supply shall go into stop mode when the IC's temperature is over its trigger point .When IC temperature is lower than regulation value the power supply shall be self-recovering.

    Short circuit protection

    The power supply shall withstand continuous current of short circuit on output no any damage. When the fault condition is removed the power supply shall be self-recovering.

    ENVIRONMENTAL REQUIREMENTS

    OPERATING CONDITION

    The power supply shall be compliant with each item in this specification for the following environmental conditions.

    Ambient Temperature

    -10℃~40℃

    Relative Humidity

    20%~90%

    STORAGE CONDITION

    Ambient Temperature

    -20℃~80℃

    Relative Humidity

    10%~90%

    Protection Function

    MTBF

    The power supply is required to have a minimum predicted MTBF (MIL-STD-217F) of 100,000 hours under normal operating conditions and typical usage.

    WEIGHT

    The weight of the power supply shall be about 120g.

    SAFETY STANDARD

    The power supply is designed for indoor use and is compliant with IEC 60601-1 and EN 60601-1 standards, as well as other relevant safety regulations.

    Electromagnetic Compatiblity

    Number

    Item

    Specifications

    Class

    Standards

    1

    CE

    CLASS B

    /

    IEC/EN60601-1-2

    YY0505

    GB4824

    EN55011

    FCC Part 18

    GB4824

    2

    RE

    CLASS B

    /

    IEC/EN60601-1-2

    YY0505

    GB4824

    EN55011

    FCC Part 18

    3

    SURGE

    Line to line±1KV

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-5

    GB17626.5

     

     

    Line to GND±2KV

    A

     

    4

    ESD

    Air discharge±15KV

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-2

    GB17626.2

     

     

    Contact discharge±8KV

    A

     

    5

    EFT/B

    ±2KV (BURST FREQUENCY=100KHz)

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-4

    GB17626.4

    6

    DIP

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-11

    GB17626.11

     

     

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

     

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

     

     

    Dropped to 0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    RS

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-3

    GB17626.3

    8

    CS

    Test frequency:0.15MHz~80MHz;

    Field intensity:6V;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-6

    GB17626.6

    9

    THD

    CLASS A (in system)

    /

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-3-2

    GB17625.1

    10

    Voltage

    Fluctuation and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-3-3

    GB17625.2

    11

    Power Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-8

    GB17626.8

    Performance criterion A: Normal performance within limits in this specification. 
    Performance criterion B: Temporary loss of some functions or degradation of performance. The recovering performance of which without operator intervention.
    Performance Criterion C: Temporary loss of function or performance degradation that requires operator intervention for recovery.
    Performance Criterion D: Loss of function or performance degradation that is not recoverable due to hardware or software damage, or data loss.

    Safety: accord with

    The power supply is designed for indoor use to meet IEC 60601-1 and EN 60601-1standards and for the following safety regulations.
     

    ITEM

    COUNTRY

    STANDARD

    UL

    USA

    UL60950-1/UL60601-1

    CE

    EUROPE

    EN60950-1/EN60601-1

    CB

    WORLDWIDE

    IEC 60601-1

    TUV

    GERMANY

    IEC60601-1

    NRTL

    CUSA

    IEC60601-1/UL60601-1

    GS

    GERMANY

    EN60601-1

    BS

    ENGLAND

    EN60601-1

    SAA

    AUSTRALIA

    AS/NZS6-1

    The primary to secondary: 4000Vac 10mA for 60 seconds

    Mechanical Requirement

    AC Plug: AUS L*W*H=62.0*43.5*50.0mm
    Acctiv (1)szr
    AC Plug:CN   L*W*H=62.0*40.5*41.0mm
    Acctiv (2)mx0
    AC Plug:US   L*W*H=62.0*40.5*41.0mm
    Acctiv (3)pgx
    AC Plug:UK   L*W*H=62.0*50.5*53.0mm
    Acctiv (4)jzl
    AC Plug:US   L*W*H=62.0*40.5*41.0mm
    Acctiv (5)ber

    DC PLUG FIGURE(Detail information selection from file of LXC_Dcline_Library.xls)

    J@KWE%C2%D~PB3~M(7GYPEFtn2
    Note: Details are selected from the LXC_ Wire Library drawing XLS file

    Label

    )UDQVJE70HM`K3AW3GHA5`38yd

    about us

    Our company introduced MES system in 2018, and built 2 fully automated production lines in 2019, which greatly improved product quality, production efficiency and capacity; At present, our factory covers an area of more than 8000 square meters, and our employees more than 300 people; It has 4 automatic placement machines, 4 dispensing machines, 4 automatic solder paste machines, 4 reflow soldering machines, 4 automatic AOI machines, 1 first piece detector, 2 automatic AI insertion machines, 4 wave soldering machines, 8 automatic production lines, 3 automatic radium carving machines, 15 ATE test systems, 10 automatic aging cabinets. In order to achieve comprehensive control, Dragon Star is equipped with various control systems such as WMS, SMT error prevention, ESOP, abnormal call, MES, ERP, etc., so that all stages of production are in the control process.