20W medical power supply module 9V 12V 15V 24V open frame reliable switching power supply ACMS25E

The power supply design for medical equipments and rang of input voltage acoording to global power grid.

The functions of protection are output current limited shorted and overvoltage etc;

Lower ripple wave and spike noise.

    Input Characteristics

    IEC68-2-1 IEC68-2-1A GB2423.1-1989 

    Basic environmental testing procedures for electric and electronic products - Test A: Low temperature test method;

    IEC68-2-2 IEC68-2-2A GB2423.2-1989 

    Basic environmental testing procedures for electric and electronic products - Test B: Dry hear;

    IEC68-2-3 GB2423.9-1989 

    Basic environmental test procedure for electrical and electronic products Test Cd: Damp heat, steady state, primarily for equipment

    IEC68-2-6 GB2423.10-1995 

    Basic environmental testing procedures for electric and electronic products Test Fc and guidance:Vibration sinusoidal

    IEC60601-1 GB 9706.1—2007

    Medical electrical equipment - part 1: general requirements for safety;

    IEC60601-1-2 YY0505-2012

    Medical electrical equipment - part 1-2: common safety requirements and standards: electromagnetic compatibility requirements and tests.

    Ambient condition

    Environment of work condition

    Number

    Item

    Specifications

    Unit

    Note

    1

    Work temp

    0~+40

     

    2

    Relative humidity

    1090

    %

    Ncondensation

    3

    Atmospheric pressure

    70K~106K

    Pa

    Normal

    4

    Height

    ≤5000

    m

    Work temp descend 1℃ every 300m over 5000m

    5

    Cooling Mode

    Free air convection

     

     

    Environment of storage condition

    Number

    Item

    Specifications

    Unit

    Note

    1

    Storage temp

    -20~+80

    No work

    2

    Storage humidity

    10~90

    %

    No condensation

    Environment of test condition

    Number

    Item

    Specifications

    Note

    1

    High temp

    +40℃

    Work normal and performance good

    2

    Low temp

    0℃

    Work normal and performance good

    3

    Storage in High temp

    +80℃

    Work normal and performance good after 2 hours in normal Environment

    4

    Storage in low temp

    -20℃

    Work normal and performance good after 2 hours in normal Environment

    5

    Vibration

    The sample can afford for 5~500HZ free vibration in X Y Z three direction. spectral density is 10 m2/sin range of 5~10Hz,3m2/sin range of 10~200HZ,1 m2/sin range of 200~500HZ.20min every direction.

    normal

    6

    Shock

    Half sine wave,peak accelerated speed 20G,last 11mS,6faces,3times

    normal

    Electromagnetic Compatiblity

    Numbe

    Item

    Specifications

    Class

    Standards

    1

    (CE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    2

    (RE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    3

    (SURGE)

    Line to line±1KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-5; GB17626.5

     

     

    Line to GND±2KV

    A

     

    4

    (ESD)

    Air discharge±15KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-2; GB17626.2

     

     

    Contact discharge±8KV

    A

     

    5

    (EFT/B)

    ±2KV (BURST FREQUENCY=100KHZ)

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-4; GB17626.4

    6

    (DIP)

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-11; GB17626.11

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

    Dropped to0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    (RS)

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;80%AM(1KHz)

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-3; GB17626.3

    8

    (CS)

    Test frequency:0.15MHz~80MHz;

    Field intensity:6Vrms;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-6; GB17626.6

    9

    (THD)

    CLASSA(in system)

    /

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-3-2; GB17625.1

    10

    Voltage

    Fluctuation and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-3-3; GB17625.2

    11

    Power Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-8; GB17626.8

    The power supply  can  be operated normaly.
    The power supply temporary decline or loss of function or performance  but free Recoverability. 
    The power supply temporary decline or loss of function or performance  but forced Recoverability.
    The power supply loss of function or performance forever.

    Safety: accord with

    The power supply is designed for indoor use to meet IEC 60601-1 and EN 60601-1standards and for the following safety regulations.

    Number

    Item

    Specifications

    1

    Fault impact 

    The power supply will left in fault and no harmful to anthing else.

    2

    Components

    The components in the device above industrial leve

    3

    Audible noise

    45dB(A),distance 1m

    4

    Cooling Mode

    Free air convection

    Outline

    Length x width x height=70.1mm * 48.5mm * 22.5Max (tolerance: ± 0.4mm)
    sdz8k

    Cable

    sda5u