15W longxc SMPS AC DC 24V 15W Open Frame Medical Switching Power Supply ACMS13

The purpose of the document is to specify the functional requirements of the ACMS13 series switching power supply

    Input Characteristics

    Input Voltage:

    Nominal Voltage: 100~240Vac

    Variation Range: 90~264Vac

    Input Frequency:

    Nominal Frequency: 50/60Hz.

    Variation Frequency: 47~63Hz

    Input Current:

    1.0Amps max at any input voltage and rated DC output and rated load.

    Inrush Current:

    60Amps max. cold start at 264Vac input, with rated load and 25℃ ambient.

    AC Leakage Current:

    Normal 0.5mA max.at 264Vac input.

    Single fault 1.0mA max.at 264Vac input.

    Output Characteristics

    Model number

    Output voltage(V)

    Output Range(V)

    Output current (A)

    Output power (W)

    Efficiency(%)

    Note

    ACMS13-050

    5.0

    4.7-5.3

    0-1.70

    8.50

    ≥80

    @220Vac

     

    ACMS13-075

    7.5

    7.2-7.8

    0-1.70

    12.75

    ACMS13-090

    9.0

    8.7-9.3

    0-1.70

    15.30

    ACMS13-120

    12.0

    11.7-12.3

    0-1.70

    20.40

    ACMS13-136

    13.6

    13.5-13.9

    0-1.47

    20.00

    ≥82

    ACMS13-142

    14.2

    14.0-14.5

    0-1.40

    20.00

    ACMS13-150

    15.0

    14.7-15.3

    0-1.33

    20.00

    ACMS13-168

    16.8

    16.5-17.1

    0-1.19

    20.00

    ACMS13-180

    18.0

    17.7-18.3

    0-1.11

    20.00

    ACMS13-190

    19.0

    18.7-19.3

    0-1.05

    20.00

    ACMS13-200

    20.0

    19.7-20.3

    0-1.00

    20.00

    ACMS13-240

    24.0

    23.5-24.5

    0-0.85

    20.40

    Line regulation

    ±1%

    Load regulation

    ±2%

    Protection Function

    Short circuit test

    The power supply will be automaticity recovery when short circuit faults remove.

    Over current Protection

    When the output current reaches 110%~200% of the rated current, the power supply will be protected. After the overcurrent fault is removed, the power supply will be automatically recovery.

    Over Voltage Protection

    When the output voltage reaches 105%~125% of the rated voltage, the power supply will be protected and can resume normal work after the fault removed.

    Over temperature protection

    5mS Min.at 115Vac input and output rated Load.

    The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering.

    Environmental Requiremet

    Operating Temperature

    0℃ to 40℃,Rated load, Normal operation.

    Storage Temperature:-20℃ to 80℃

    No work

    Storage Humidity: 10%~90%

    No condensation

    Atmospheric pressure

    70-106KPa, Normal.

    Altitude

    5000m,Work temp descend 1℃ every 300m over 5000m.

    (9~200Hz,Acceleration 5m/S2)

    Transportation: 5-9Hz, A=3.5mm

    Acceleration=5m/S2

    Acceleration=15m/S2

    Axes, 10 cycles per axis

    No permanent damage may occur during testing.

    The Power supply can be restore to normal after power off/on.

    Dropping Packed

    1m for wall mount type and 760mm for desktop type as above described.

    The horizontal surface consists of hardwood at least 13mm thick, mounted on two layers of  plywood and away  from  land 19mm ~ 20mm.

    Relative Humidity

    5%(0℃) ~90%(40℃)RH, 72Hrs, Rated load, Normal operating.

    MTBF

    The power supply shall have a minimum predicted MTBF(MIL-STD-217F) of 100000 hours at normal operation condition and normal use.

    Electromagnetic Compatiblity

    Numbe

    Item

    Specifications

    Class

    Standards

    1

    (CE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    2

    (RE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    3

    (SURGE)

    Line to line±1KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-5; GB17626.5

     

     

    Line to GND±2KV

    A

     

    4

    (ESD)

    Air discharge±15KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-2; GB17626.2

     

     

    Contact discharge±8KV

    A

     

    5

    (EFT/B)

    ±2KV (BURST FREQUENCY=100KHZ)

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-4; GB17626.4

    6

    (DIP)

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-11; GB17626.11

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

    Dropped to0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    (RS)

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;80%AM(1KHz)

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-3; GB17626.3

    8

    (CS)

    Test frequency:0.15MHz~80MHz;

    Field intensity:6Vrms;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-6; GB17626.6

    9

    (THD)

    CLASSA(in system)

    /

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-3-2; GB17625.1

    10

    Voltage

    Fluctuation and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-3-3; GB17625.2

    11

    Power Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-8; GB17626.8

    Performance criterion A: Normal performance within limits in this specification. 
    Performance criterion B: Temporary loss of some functions or degradation of performance. The recovering performance of which without operator intervention.
    Performance criterion C: Temporary loss of function or degradation of performance .The recovering performance of which requires operator intervention.
    Performance criterion D: Loss of function or degradation of performance which is not recoverable .Owing to damage of hardware software, or loss of data.

    Safety: accord with

    The power supply is designed for indoor use to meet IEC 60601-1 and EN 60601-1standards and for the following safety regulations.
     

    Item

    Country

    Standard

    UL

    USA

    UL60950-1/UL60601-1

    CE

    Europe

    EN60950-1/EN60601-1

    CB

    Worldwide

    IEC60601-1

    TUV

    Germany

    IEC60601-1

    NRTL

    CUSA

    IEC60601-1/UL60601-1

    GS

    Germany

    EN60601-1

    BS

    England

    EN60601-1

    SAA

    Australia

    AS/NZS6-1

    The primary to secondary: 4000Vac 10mA for 60 seconds

    Outline

    The power supply size: L 49.0x W 23.8 x H 22.0 mm(Max);